Hebei Aixuanda medical equipment Co.,Ltd
|
Place of Origin: | Zhejiang, China (Mainland) |
Add to My Favorites | |
HiSupplier Escrow |
Automatic Medical Vacuum Pumps , Single Channel Syringe Pump
Quick Detail:
174 Drugs with 10 Categories of Drug Library
Description:
Features:
1. Four working modes:
Rate mode, Time mode, Weight mode and Drug mode
2. 174 Drugs with 10 Categories of Drug Library
Cardiovascular Sys, Respiratory Sys, Antibiotics Sys, Alimentary Sys.
Nervous Sys, Analgesic Sys, Urinary Sys, Blood Sys, Endocrine Sys.
17 units of dose: Ug/kg/min, mg/kg/min, g/kg/min, ug/kg/h, mg/kg/h, g/kg/h, ug/min, mg/min, g/min, ug/h, mg/h, g/h, U/min, IU/h, U/kg/h, ng/kg/min, can make drug dispensing more convenient and safer.
3. ICU infusion workstation: Combines for syringe pumps and one infusion
Pump with Infusion Support into one ICU Infusion Workstation, sharing a Power Interface.
4. Stackable up to the maximum 8 channels. Transfer the message between the Nurse-call System and the pumps.
5. Bolus Function.
Specifications:
1. Compatible Syringe: Any standard syringe with 5ml, 10ml, 20ml, 30ml, 50ml and 60ml.
2. Accuracy: ± 2% (Any manufacturers of syringes, right after calibration)
Delivery rate range:
50 ml syringe - 0.1 1800 ml / hr (0.1 ml / hr step)
30 ml syringe - 0.1 1200 ml / hr (0.1 ml / hr step)
20 ml syringe - 0.1 800 ml / hr (0.1 ml / hr step)
10 ml syringe - 0.1 400 ml / hr (0.1 ml / hr step)
5 ml syringe - 0.1 200 ml / hr (0.1 ml / hr step)
3. Time setting mode range:
Volume: 0.1 999.9ml, 0.1ml step.
Time: 0.1 999.9min, 0.1min step.
4. Drug Library setting mode range:
Sol. Volume: 0.1 199.9ml, 0.1ml step.
Drug mass: 0.1 999.9ml, 0.1 ml step.
Dose: 0.001 9999, 0.001 step.
5. Bolus setting mode range:
Bolus. Volume: 0.1ml 60.0ml, 0.1ml step
Bolus Rate: 5ml: 0.1ml/h 200.0ml/h, 0.1ml/h step
10ml: 0.1ml/h 400.0ml/h, 0.1ml/h step
20ml: 0.1ml/h 800.0ml/h, 0.1ml/h step
30ml: 0.1ml/h 1200.0ml/h, 0.1ml/h step
50ml(60ml): 0.1ml/h 1800.0ml/h, 0.1ml/h step
6. Purge 50ml(60ml): 1800.0ml/h, 20ml(30ml); 800.0ml/h
10ml: 400.0ml/h
5ml: 200.0ml/h
7. KVO rate: 0.1ml/h-1.0ml/h, 0.1ml/h step
8. Alarms: Near empty, Syringe empty, Volume complete (a volume limit has been set), Occlusion, Syringe loss, Syringe dislocated, Operation error, Low battery and other device failures.
9. Power supply: AC 100-240V +/- 10%, 50/60Hz. 10. Power consumption: 25<=VA
10. Internal Lithium Battery: 11.1V, 2000mAh, Approximately 8 hours of continuous operation in a new and fully charged battery running at 5ml/h.
11. Safety level: Class I, Type: CF
12. Size: 230mm(L) x 150mm(W) x 125mm(H)
13. Net Weight: 2.2Kg. (Approx)
14. Operating conditions:
Ambient temperature: 5-40° C Relative humidity: 20 to 90%
15. Storage conditions:
Ambient temperature: -10-55° C Relative humidity: 10 to 90%
16. Accessories: AC power line, Operational manual
syringe pump(JAZ-6T) | |
The specification of syringe pump | |
Flow rate | 0.1ml/h-1500ml/h(0.1ml/h increments, 1ml/h increments when rate is over 1000ml/h )50ml syringe |
0.1ml/h-900ml/h (0.1ml/h increments)30ml syringe | |
0.1ml/h-600ml/h(0.1ml/h increments)20ml syringe | |
0.1ml/h-400ml/h(0.1ml/h increments)10ml syringe | |
Bolus rate | 1500ml/h(50 syringe); 900ml/h(30ml syringe); 600ml/h(20ml syringe); 400ml/h(10ml) |
Occlusion pressure | 40N, 60N(± 10N) |
Volumetric accuracy | ≤ 2%(drive accuracy ≤ 1%) |
Total volume indication | 0.1-9999ml |
Preset limited volume | 0.1-9999ml |
Alarm | Near empty, Empty, Syringe dislocated, Syringe plunger disengaged, Malfunction, Forget to operate after power on, Occlusion, Low battery, Battery volume exhausted, AC power failure, Preset limited volume is finished |
Operating condition | Temperature: -40 to 40degree; Relative humdity: 20 to 90% |
Transport and | Temperature: -40 to 55degree |
storage condition | |
External dimensions | 306(L)× 135(W) × 127(H)mm |
Net weight |